Bo Ling Instrument teaches you how to make the centrifuge last longer

Generally, the speed of the refrigerated centrifuge with refrigeration is not too high, and the minimum temperature that can be reached is about -20 to -9 degrees. The centrifuge is mostly used to collect microorganisms, cell debris, cells, large organelles, sulfate precipitates, and immunoprecipitates Wait. In order to ensure the speed of the refrigerated centrifuge, the following should be familiar:

1. After centrifugation, wipe off the water in the centrifugal cavity. Apply a little neutral lubricating grease to the conical surface of the motor shaft every week to prevent corrosion of the rotating shaft. If the centrifuge is not used for a long time, the rotor should be taken out, wiped clean and placed in a dry place to prevent corrosion.

2. The main power plug should be removed when the instrument is not used for a long time or when it is being repaired. Otherwise, the instrument will be electrified, especially safety accidents are likely to occur during maintenance.

3. In order to protect the refrigerating compressor, the interval between power off and power on must be greater than 3 minutes, otherwise the compressor will be damaged.

4. When the rotor is not in use, it should be taken out of the centrifugal cavity, cleaned and wiped dry with neutral washing liquid in time to prevent chemical corrosion, and stored in a dry and ventilated place. It is not allowed to scrub the rotor with non-neutral detergent, and it is not allowed to blow (dry) the rotor with electric hot air. The center hole of the rotor should be coated with a little grease.

5. In order to ensure the freezing effect, when the ambient temperature is higher than 30 ℃, the rotor and the centrifugal cavity should be pre-cooled, and the rotor should also be operated at a reduced speed of 15%.

6. Centrifuge tubes should be updated regularly, and the use of centrifuge tubes on the verge of rupture is strictly prohibited.

7. Polypropylene (PP) centrifuge tubes (bottles) cannot be used in contact with gasoline, kerosene, etc. Polycarbonate (PC) centrifuge tubes (bottles) cannot be used in contact with gasoline, acetaldehyde, acetone, ethanol, isobutanol, cresol, etc. Polyethylene (PE) centrifuge tubes (bottles) cannot be used in contact with sulfuric acid (50%, 75%), benzene, gasoline, kerosene, etc.

8. Before each use, pay attention to check the rotor for corrosion points and fine cracks. It is forbidden to use corroded or cracked rotors, and use rotors that exceed the shelf life to ensure personal safety.

9. When using the centrifuge rotor, make sure that the set rotor number is correct. If the rotor number is set incorrectly. Will cause the rotor to be used at overspeed or fail to achieve the required centrifugal effect. In particular, the malignant accidents that may cause the rotor to burst when used at overspeed must not be negligent.

Multipurpose Diaper Backpack are becoming more and more popular as regular items for babies and parents. Whether it is a long-distance travel or a short trip, Diaper Bag provides important convenience for babies and parents. The popularity of Diaper Bags has caused new parents to pay more attention to infant travel, not only psychologically but also physical changes, with great significance.