Constant Ord's new petroleum wax oil content tester technical parameters


Petroleum wax oil content tester / paraffin oil content tester model; HA3554

I. Use and scope of application
This instrument is designed and manufactured according to the requirements of the People's Republic of China standard GB/T3554 "Measurement method for petroleum wax oil content", and is suitable for the determination of petroleum wax with a freezing point above 30 °C and a oil content of not more than 15%. Oil content.
The main points of the method are: dissolving the sample in methyl ethyl ketone, cooling to -32 ° C, separating the wax, filtering and distilling off the butanone in the filtrate, weighing the residual oil, and calculating the oil content of the wax.
Note: Some waxes with a oil content greater than 5% should not be used if the oil is not completely dissolved in methyl ethyl ketone.
Instrument characteristics
1. The -35 ° C cold bath is an independently designed refrigeration unit. There are 4 cold tanks, which can be tested in groups.
2, wax bath, 35 ° C evaporator designed as a whole, the lower half is room temperature ~ 90 ° C wax bath, the temperature of the water bath is controllable, adjustable; the upper part is the air bath evaporation device, the temperature is controlled 35 ° C.
3, cold bath, wax bath, evaporator temperature adjustment and control are digital temperature controller, temperature adjustment and control is accurate and intuitive.
4. The amount of blown air entering the sample in the evaporation device is detected by the glass rotor flowmeter, and the blow volume can be accurately controlled according to the specific requirements of the test. The blow time can be arbitrarily set by the timer within 60 minutes according to the test time requirement.
5. The overall design of the instrument is a desktop structure with novel design, reasonable structure and convenient use.
Main technical characteristics and parameters
1, working power: AC (220 ± 10%) V, 50Hz.
2, wax bath temperature: room temperature ~ 90 ° C ± 5 ° C.
3. Cold bath temperature: -35 °C ± 1 °C.
4. Evaporation temperature: 35 °C ± 1 °C.
5, the total flow of air blowing: (8 ~ 12) L / min (4 nozzles total gas output).
6. Timer timing range: (0~60) min.
7, cold bath dimensions: 510mm × 400mm × 510mm (length × width × height).
8. Water bath evaporator Dimensions: 380mm × 320mm × 530mm (length × width × height).
9, instrument power consumption: cold bath: no more than 2000W; water bath evaporator: no more than 800W.